SensoLite Nova
If 10 test measurements are done after the first appearance of the
battery icon, the error message "E-6" will be displayed on the
screen when the meter is switched on. It means that it is not
allowed to perform any more tests until the batteries are changed.
Important! Keep meter switched off while replacing
batteries.
The battery holder is located on the back side of the meter.
Pull battery holder to the side and take it apart from the
device. Replace the old, discharged batteries with new ones.
Make sure the ”+” and ”” poles are properly positioned.
Click the battery cover back into its place, then check the
equipment’s ability to operate by switching the meter on.