SensoLite Nova
Meter is damaged. Check the meter with Check-strip, see “SensoLite Nova
system checking” chapter.
7. E-6 error message
Batteries are low. Replace the batteries.
High intensity electrostatic discharge may influence the meter. Please remove the
batteries and keep the OK button pressed about 15-20 seconds. Replace the batteries
and turn on the meter. If E-6 appears again, please change the batteries.
8. E-7 error message
Insufficient amount of blood was applied on strip. Repeat test with bigger drop of
blood.
High intensity electrostatic discharge may influence the meter. Please repeat the
measurement.
9. E-8 error message
Test wasn't performed properly. Repeat test with new test strip and be more careful
when performing each step.